Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 88: (1 year) CAL VER /5500 DATE: 27-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 57 NUMBER OF LINES: 220 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 88 Service Manual (PN 880906 Dec. 1990, Rev. 2, 9/94). # # Compatibility: # 5500/CAL or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55_88_2W.bmp # 55_88_CW.bmp # 55_88_LA.bmp # 55_88_HA.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P J F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 ASK- U 1.007 HEAD {BASIC METER OPERATIONS} 1.008 DISP Turn the Meter rotary switch to Ohms. 1.009 DISP Connect a test lead from V Ohm Diode input to the 1.009 DISP mA input. (If you are using a test probe, touch 1.009 DISP the half of the input contact nearest the LCD.) 1.010 ACC 1.000kZ TOL 1.011 MEMI Enter UUT reading in kilohms: 1.012 MEME 1.013 MEMC kZ 0.005U 2.001 DISP Remove the test lead from the UUT. 2.002 HEAD {15A FUSE TEST} 2.003 DISP Insert the plug end of the test lead into the A input. 2.004 EVAL Is the UUT beeping repeatedly? 3.001 HEAD {1A FUSE TEST} 3.002 DISP Insert the plug end of the test lead into the mA input. 3.003 EVAL Is the UUT beeping repeatedly? 4.001 DISP Remove the test lead. 4.002 HEAD {DISPLAY TEST} 4.003 DISP With the Meter rotary switch in the OFF position, press 4.003 DISP and hold down any button and turn the rotary switch 4.003 DISP to any function setting. 4.004 EVAL Did the UUT display test pass? 5.001 HEAD {ROTARY SWITCH TESTS} 5.002 DISP Press and hold down RPM button for 2 seconds while 5.002 DISP rotating the UUT function switch from OFF to {VAC}. 5.003 MATH MEM1 = 0 5.004 MEMI Enter UUT reading: 5.005 MEME 5.006 MEMC display 12U 6.001 DISP Rotate the UUT function switch to {VDC}. 6.002 MATH MEM1 = -32 6.003 MEMI Enter UUT reading: 6.004 MEME 6.005 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {mVDC}. 7.002 MATH MEM1 = -64 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 DISP Rotate the UUT function switch to {Ohms}. 8.002 MATH MEM1 = -96 8.003 MEMI Enter UUT reading: 8.004 MEME 8.005 MEMC display 12U 9.001 DISP Rotate the UUT function switch to {Diode}. 9.002 MATH MEM1 = -128 9.003 MEMI Enter UUT reading: 9.004 MEME 9.005 MEMC display 12U 10.001 DISP Rotate the UUT function switch to {mA/A DC}. 10.002 MATH MEM1 = -160 10.003 MEMI Enter UUT reading: 10.004 MEME 10.005 MEMC display 12U 11.001 DISP Rotate the UUT function switch to {mA/A AC}. 11.002 MATH MEM1 = -192 11.003 MEMI Enter UUT reading: 11.004 MEME 11.005 MEMC display 12U 12.001 ASK+ U 12.002 HEAD {AC VOLTAGE TESTS} 12.003 DISP Select Autorange VAC by pressing and holding the RANGE 12.003 DISP button for 2 seconds while rotating the UUT function 12.003 DISP switch from OFF to the VAC setting. 12.004 PIC 55_88_2W 12.005 HEAD AC VOLTAGE TESTS: {400mV Range} 12.006 5500 400 350.0mV 6.3U 60H SI 2W 13.001 5500 400 350.0mV 6.3U 1kH SI 2W 14.001 5500 400 350.0mV 7.7U 5kH SI 2W 15.001 HEAD AC VOLTAGE TESTS: {4V Range} 15.002 5500 4 3.500V 0.058U 60H SI 2W 16.001 5500 4 3.500V 0.058U 1kH SI 2W 17.001 5500 4 3.500V 0.072U 5kH SI 2W 18.001 HEAD AC VOLTAGE TESTS: {40V Range} 18.002 5500 40 35.00V 0.58U 60H SI 2W 19.001 5500 40 35.00V 0.58U 1kH SI 2W 20.001 5500 40 35.00V 0.72U 5kH SI 2W 21.001 HEAD AC VOLTAGE TESTS: {400V Range} 21.002 5500 400 350.0V 5.8U 60H SI 2W 22.001 5500 400 350.0V 5.8U 1kH SI 2W 23.001 5500 400 350.0V 7.2U 5kH SI 2W 24.001 HEAD AC VOLTAGE TESTS: {1000V Range} 24.002 5500 1000 900V 28U 60H SI 2W 25.001 5500 1000 900V 28U 5kH SI 2W 26.001 HEAD {FREQUENCY TESTS} 26.002 DISP Press RANGE to lock the 400mV Range. 26.002 DISP Press RPM three times to choose the Frequency mode. 26.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 26.004 5500 20 19.000kH 0.003U 150mV SI 2W 27.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 27.002 5500 200 190.00kH 0.03U 150mV SI 2W 28.001 HEAD {FREQUENCY SENSITIVITY AND TRIGGER LEVEL} 28.002 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V AC Range} 28.003 DISP Press the RANGE button to lock the 4V Range. 28.004 5500 1000.0H 0.2U 300mV SI 2W 29.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V DC Range} 29.002 DISP Rotate the UUT function switch to VDC. 29.002 DISP Press RANGE select the 4V Range. 29.002 DISP Press RPM three times to choose the Frequency mode. 29.003 5500 1000.0H 0.2U 1.7V SI 2W 30.001 5500 1000.0H 1.0V SI S 2W 30.002 EVAL Is the UUT {displaying all zeros}? 31.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {40V DC Range} 31.002 DISP Press the RANGE button to lock the 40V Range. 31.003 5500 1000.0H 0.2U 6.0V SI 2W 32.001 5500 1000.0H 2.0V SI S 2W 32.002 EVAL Is the UUT {displaying all zeros}? 33.001 HEAD {DC VOLTAGE TESTS} 33.002 DISP Press RPM to select VDC. 33.002 DISP 33.002 DISP Press the RANGE for 2s to select autorange. 33.003 HEAD DC VOLTAGE TESTS: {4V Range} 33.004 5500 4 3.500V 0.007U 2W 34.001 HEAD DC VOLTAGE TESTS: {40V Range} 34.002 5500 40 35.00V 0.07U 2W 35.001 5500 40 -35.00V 0.07U 2W 36.001 HEAD DC VOLTAGE TESTS: {400V Range} 36.002 5500 400 350.0V 0.7U 2W 37.001 HEAD DC VOLTAGE TESTS: {1000V Range} 37.002 5500 1000 1000V 5U 2W 38.001 HEAD {DC MILLIVOLT TEST} 38.002 DISP Rotate the UUT function switch to mVDC. 38.003 HEAD DC MILLIVOLT TEST: {400mV Range} 38.004 5500 400 350.0mV 0.7U 2W 39.001 HEAD {DIODE TEST} 39.002 DISP Rotate the UUT function switch to Diode. 39.003 5500 3.000V 0.061U 2W 40.001 HEAD {RESISTANCE TESTS} 40.002 DISP Rotate the UUT function switch to Ohms. 40.003 HEAD RESISTANCE TESTS: {400 MOhm Range} 40.004 DISP Press the Range button twice for 400 MOhm range. 40.005 5500 400 100MZ 3U 2W 41.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 41.002 DISP Press the RANGE button for 2s to select autorange. 41.003 5500 40 19.00MZ 0.22U 2W 42.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 42.002 5500 4 1.900MZ 0.005U 2W 43.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 43.002 PIC 55_88_CW 43.003 5500 40 19.00kZ 0.05U CW 44.001 HEAD RESISTANCE TESTS: {400 Ohm Range} 44.002 5500 400 190.0Z 0.5U CW 45.001 HEAD {DC MILLIAMP TESTS} 45.002 DISP Rotate the UUT function switch to mA/A DC. 45.003 PIC 55_88_LA 45.004 HEAD DC MILLIAMP TESTS: {40mA Range} 45.005 5500 40 35.00mA 0.30U 2W 46.001 HEAD DC MILLIAMP TESTS: {400mA Range} 46.002 5500 400 350.0mA 3.0U 2W 47.001 HEAD {AC MILLIAMP TESTS} 47.002 DISP Rotate the UUT function switch to mA/A AC. 47.003 HEAD AC MILLIAMP TESTS: {40mA Range} 47.004 5500 40 35.00mA 1.15U 60H SI 2W 48.001 5500 400 35.00mA 1.15U 1kH SI 2W 49.001 HEAD AC MILLIAMP TESTS: {400mA Range} 49.002 5500 400 350.0mA 5.8U 60H SI 2W 50.001 5500 400 350.0mA 5.8U 1kH SI 2W 51.001 HEAD {AC AMP TESTS} 51.002 PIC 55_88_HA 51.003 HEAD AC AMP TESTS: {4000mA Range} 51.004 5500 4000 3500mA 58U 60H SI 2W 52.001 5500 4000 3500mA 58U 1kH SI 2W 53.001 HEAD AC AMP TESTS: {10A Range} 53.002 5500 10 10.00A 0.20U 60H SI 2W 54.001 5500 10 10.00A 0.20U 1kH SI 2W 55.001 HEAD {DC AMP TESTS} 55.002 DISP Rotate the UUT function switch to mA/A DC. 55.003 HEAD DC AMP TESTS: {4000mA Range} 55.004 5500 4000 3500mA 30U 2W 56.001 HEAD DC AMP TESTS: {10A Range} 56.002 5500 10 10.00A 0.15U 2W 57.001 END